History of GE (No. 12)


FrontPage

https://cni.stanford.edu/wiki/Main_Page

ASSETArray Spatial Sensitivity Encoding TechniquesSENSE
ARCAuto calibrating Reconstruction for Cartesian imagingGRAPPA
PUREPhased array Uniformity EnhancementPrescan Normalize
SCICSurface Coil Intensity Correction
SCENICITK N4
Hyper SENSECompressed sensing

https://mri-q.com/uploads/3/4/5/7/34572113/ge_fieldnotes_volume1-2_coils.pdf

Linear(3) + HO(9) shims

x,y,z,zy,x2y2,zx,xy,z2,z3,z2x,z2y,zx2y2

readallcurrent

#####################################################
Read current linear shims
#####################################################

Please Check the Service Key. It is not present or expired.

Host name: GEMR

Mon Apr 17 15:01:31 2023
-> Command to svat: PSC_SAVE_SHIM 
Mon Apr 17 15:01:31 2023
-> Command to svat: SLEEP .2 

current values in linear shims
x = "-41" y = "-148" z = "116"

done
######################################################
Z2 at -564 mAmps
ZX at -328 mAmps
ZY at 573 mAmps
X2Y2 at -139 mAmps
XY at -353 mAmps
Z3 at 184 mAmps
Z2X at -101 mAmps
Z2Y at -486 mAmps
ZX2Y2 at 0 mAmps

Setting Shim value

> setNavShimCurrent ZX2Y2 -4000

Linear shim data file(save/load)

/w/config/mr_gshim_config

CVs

Name
autolockpfile save flag
rhfermrfermi filter application0/1(on/off)
rhrcxres
rhrcyres
rhimsize

After TPS reset button on

>ice_term&

Dicom Tags

tagsinfoexample
0043 10B2table position
0043 108Eline width/ws level15,0,99

sage<RET>

load raw data(pfile)

Log files /usr/g/service/log

gesys_GEMR : log messages at the bottom
MrMain : log at the right upper messages

Protocol file(pdf)

1.ProtoCopy 
2.ProtocolExchange>Export Mode>drag the set to Protocol Selection section>Preview
3./usr/g/service/log/Exp_Protocols/GE_ProtocolExchange/***.html ***.pdf
You should copy the files otherwise you lose them.