History of XA60upgrade (No. 10)


FrontPage

Problems

Bootup issues

 Occasional fail during the boot process

Sudden uncontrollable software during manual shim

Unwilled experimental termination made automatic table movement.

ideacmdtool IMA transfer

MRS data cannot be transferred automatically.

Cannot run with an error

CMRR-C2P ShimCache
FASTESTMAP

System log file transfer

Second decimal place cannot be entered.

 Slice thickness parameter

MREG

Recon?

command line issue

C:\ :P <RET> Cause error
Error: the system cannot find the device specified.

Info

/System/Adjustment/b0map has gradient shim types as protocols.
Is raidtool useful for us?

Host

Windows10 LoT Enterprise LTSC
128GB RAM
1TB SSD
12th Gen Intel(R) Core(TM) i9-12900 2.40GHz
16bit DICOM save(iceConfig.env)

Mars

Debian GNU/Linux 11(bullseye)
Linux mars 5.10.89-xenomai-l #1 SMP IRQPIPE Wed Jan 26 12:20:57 UTC 2022 x86_64 GNU/Linux
/dev/nvme0nl: 1.75TiB
SAMSUNG MZ1L21T9HCLS-00A07
Mem: 251Gi
32 processors?
Intel(R) Xeon(R) Gold 6226R CPU @ 2.9GHz
16Cores
I225-T1 10/100/1000/2.5GBASE-T (TX) 222.222.222.3
I219-LM 1000Base-T 192.168.2.1

In future

  • af command update
  • lncarts command update
  • dgs command update
  • dcmall,ppall command check
  • Yarra setup for K-space data transfer

Solved issues

Protocol conversion has some problem.

CMRR MB sequence conversion has some issues. (FatSat and others)
Sinc exc. pulse BWTP setting
 BWTP=1.0 but BWTP=5.2 for default. 1.0->5.2 fixed the slice-intensity variation problem.
Fat sat setting should be FA_fsat=110 (not 0)

ND data is saturated.

T1MPRAGE_ND is saturated which is used for MRS.
The ND calculation procedure handles data as 12bit integer. 16bit MPRAGE data turns to be saturated after the ND image calculation.
dOverallImageScaleCorrectionFactor (DICOMout factor)

Some DICOM Tag data has been changed..

Manual DICOM transfer

Cannot transfer to the subfolder of SAMBA.